A new 17b estradiol assay on the Kryptor® (Cis Bio International) : utilization to monitor ovulation stimulation |
Annales de Biologie Clinique. Volume 58, Number 1, 71-9, Janvier - Février 2000, Articles originaux
|
Résumé
Article gratuit
|
Author(s) : J. Taïeb, C. Sarnel, C. Benattar, A. Lindenbaum |
Summary : We studied the performances of the 17b-estradiol determination with the Kryptor System (Cis Bio international), in assisted medical procreation cycles. The detection is based on Trace® technology. We compared this method with Estradiol-6® method on ACS-180 (Bayer Diagnostics). The Kryptor method is sensitive, has a good precision and accuracy. Comparison with ACS-180 showed a good correlation but results were much higher. Patients under stimulation for assisted procreation were followed by the two methods. The results were close to those obtained with ACS-180 but significantly higher. Thus it was necessary to define the levels which have to be reached before ovulation induction. |
Keywords : 17b-estradiol – Kryptor – Trace® technology – Immunoassay. |
|