Texte intégral de l'article
 
 

Evaluation of the analytical performances of CRP Diasys ® reagent on Roche Hitachi 917 ®


Annales de Biologie Clinique. Volume 64, Number 4, 335-9, Juillet-Août 2006, Pratique quotidienne

Résumé   Article gratuit  

Author(s) : M Pachot, F Péronnet, C Villard, A Bayle

Summary : C reactive protein, the most sensible acute phase protein of inflammation and the labororatory should perform CRP testing on a continous 24 hour basis. The measurement is mainly performed by immunoturbimetry and immunonephelemetry methods available on multiparametric biochemical analyzer. In this study, we evaluated the analytical performances, precision and exactitude, of the CRP Diasys ® reagent on Roche Hitachi 917 ®. The results were compared to those obtained with a CRP latex immunoassay (Roche). The reagent showed high analytical characteristics and especially a significant precision in a large range of CRP levels including low levels between 1 and 3 mg/L. Although this reagent is not considered as a high-sensitive CRP reagent, the measurement quality obtained in the 1-3 mg/L range allows an utilization as a cardiovascular risk predictor.

Keywords : C-reactive protein, high-sensitivity CRP, immunoturbidimetric assay, cardiovascular risk predictor

Copyright © 2007 John Libbey Eurotext - Tous droits réservés